THERMO CAD detector with ACQUITY UPLC H-Class freezes after first injection - WKB6659
Article number: 6659
SYMPTOMS
- The first injection of a sequence is always successful
- Run Samples freezes when a second injection of the same sequence is made
- An error of Z-axis fault occurs after ACQUITY UPLC H-Class times out
ENVIRONMENT
- ACQUITY UPLC H-Class
- THERMO CAD detector
- Empower 2
- THERMO CAD drivers for Empower 2
CAUSE
Unknown
FIX or WORKAROUND
- The THERMO CAD detector has one contact closure cable going to QSM stop flow and one contact closure cable going to the FTN inject start.
- Disconnect the cable going to QSM but leave the FTN inject start cable.
- System is now able to run the entire sequence.