Pressure trace flatlines on ACQUITY UPLC BSM - WKB65852
Article number: 65852
SYMPTOMS
- Pressure trace completely drops out to a flat line at about 0 psi during pump operation
ENVIRONMENT
- ACQUITY UPLC BSM
CAUSE
Faulty transducer
FIX or WORKAROUND
Replace the faulty transducer.
ADDITIONAL INFORMATION
Ensure that there is not a mix of 15kpsi and 20kpsi transducers on the system; this situation can cause sync errors.
id65852, UPBINARY, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM