CTC PAL Autosampler pauses upon injection of second sample - WKB64649
Article number: 64649
SYMPTOMS
- CTC PAL Autosampler pauses upon injection of second sample
- First sample injects as expected
ENVIRONMENT
- CTC PAL Autosampler
CAUSE
Corrupt CTC PAL Autosampler file (.cyx)
FIX or WORKAROUND
Replace the CTC PAL Autosampler file (.cyx) with one that is known to work, or rewrite the CTC PAL Autosampler file (.cyx) from scratch.