Different chromatography on different FLRs - WKB64135

Article number: 64135

SYMPTOMS

  • Response for assay varies across different FLR systems in the lab
  • Passes Raman signal to noise and calibration
  • Gain linearity varies across different FLR systems

ENVIRONMENT

  • ACQUITY UPLC FLR Detector

CAUSE

Systems are not normalized to the same level.

FIX or WORKAROUND

  1. Run 100% water (LCMS-grade if possible) through the FLR at 0.5 mL/min for an hour with lamp on.
  2. Normalize FLR.

ADDITIONAL INFORMATION

Use the highest-quality water for calibration, normalization, and Raman signal to noise.

id64135, 2475, 2475B, A-10FL, A-30FL, UPFLRARC, UPFLRDET, UPPFLR, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO