Different chromatography on different FLRs - WKB64135
Article number: 64135
SYMPTOMS
- Response for assay varies across different FLR systems in the lab
- Passes Raman signal to noise and calibration
- Gain linearity varies across different FLR systems
ENVIRONMENT
- ACQUITY UPLC FLR Detector
CAUSE
Systems are not normalized to the same level.
FIX or WORKAROUND
- Run 100% water (LCMS-grade if possible) through the FLR at 0.5 mL/min for an hour with lamp on.
- Normalize FLR.
ADDITIONAL INFORMATION
Use the highest-quality water for calibration, normalization, and Raman signal to noise.
id64135, 2475, 2475B, A-10FL, A-30FL, UPFLRARC, UPFLRDET, UPPFLR, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO