Pressure variation and reproducibility issues on H-Class SM-FTN - WKB63186
Article number: 63186
SYMPTOMS
- Pressure fluctuation and standard reproducibility issues on H-Class
- System adequately primed
- Needle seal readiness test failure
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
The needle is damaged.

Needle drape when in retracted position; the shape indicates that needle has been pulled forward

New needle
FIX or WORKAROUND
- Replace the needle using Sample Needle Kit, 15 µL, standard (700005215).
- Run the needle seal readiness test.
ADDITIONAL INFORMATION
It is important to position and remove trays carefully because the impacts can be severe.