In-line filter leaks on 2695 - WKB62487

Article number: 62487

SYMPTOMS

  • In-line filter leaks on 2695

ENVIRONMENT

  • Alliance HPLC System

CAUSE

Damaged Inline Filter Assembly or tubing warped

FIX or WORKAROUND

Replace Inline Filter Assembly WAT035190

or 

Trim tubing to remove old ferrule and screw and replace with WAT025604

ADDITIONAL INFORMATION

Order the following parts as well as the fittings into the inline filter can be problematic:

  1. Assembly, Tubing Vent Left-In Line (# WAT270974)
  2. SS Tubing, Filter - V1 (# WAT270975)