Skip to main content
Waters

1515 "High Pressure Shutdown" error in Breeze 2 - WKB62116

Article number: 62116

SYMPTOMS

  • Running Waters Beverage Analysis with column (Waters XBridge BEH Phenyl 2.5 µm 4.6 x 75 mm) connected, a "High Pressure Shutdown" error appears in Breeze 2
  • With union connected, pressure is steady and low
  • System is flushed with 100% water
  • Multiple Waters XBridge BEH Phenyl 2.5 µm 4.6 x 75 mm columns bring the same error
  • With column connected directly to the outlet of the 1515, the same error occurs

ENVIRONMENT

  • Waters Beverage Analysis
  • 1515 Isocratic HPLC Pump
  • Breeze 2
  • 717plus Autosampler
  • 2487 Dual-Wavelength Absorbance Detector

CAUSE

Clogged column

FIX or WORKAROUND

Replace the column

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.