Instruments show up as "NO" for OK? column in Node Properties - WKB61704

Article number: 61704

SYMPTOMS

  • Instruments show up as "NO" in Node Properties
  • Unable to connect to system

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2487 Dual-Wavelength Absorbance Detector
  • Empower 3

CAUSE

 

FIX or WORKAROUND

Note: Before disconnecting any IEEE cables, ensure that there are no runs in process on any system connected to the node you will be working with.

Connect the IEEE cables as follows:

  1. Start with the last instrument.
  2. Connect the IEEE cable to it.
  3. Connect directly into the next instrument, working back to and ending with the Acquisition Server.

ADDITIONAL INFORMATION

 