System overpressure error with ACQUITY I-Class and Xevo G2 ToF - WKB6125

Article number: 6125

SYMPTOMS

  • System overpressure error after 10 minutes of running samples
  • Changing column and guard column does not correct
  • No blocks found in the pump and sample manager

ENVIRONMENT

  • ACQUITY I-Class System
  • Xevo G2 ToF

CAUSE

Blocked tube leading to ESI inlet in the MS

FIX or WORKAROUND

Replace with black PEEK tubing (700003799 - Black PEEK 1/16" x 0.004 x 60").

ADDITIONAL INFORMATION

 

id6125, Electrospray, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, XEVOG2QTOF, XEVOG2TOF