Error "Stop flow from remote host" with QDa detector - WKB6115
Article number: 6115
SYMPTOMS
- Stop flow from remote host error in the Console
- No signal when running samples
- Reboot does not resolve
- Nitrogen supply is sufficient
ENVIRONMENT
- QDa detector
- Performance style
- Classic ACQUITY
CAUSE
Bad capillary probe
No readback on cone voltage
FIX or WORKAROUND
- Replace the probe (700009642 - ESI Probe Assy, 500mm x 100µm, ACQUITY).
- Open the Tune page (available only under Service mode).
- Switch from negative to positive mode and then back to negative mode.
- Cone voltage now reads a value.
ADDITIONAL INFORMATION
id6115, Electrospray, QDA