Rotary Tray not calibrated error message on FTN Sample Manager - WKB6095

Article number: 6095

SYMPTOMS

  • Error message reads "Rotary Tray not calibrated".

ENVIRONMENT

  • ACQUITY Sample Manager FTN
  • ACQUITY M-Class Sample Manager-FL
  • ACQUITY I-Class Sample Manager-FL

CAUSE

Unknown.

FIX or WORKAROUND

To calibrate the rotary tray, perform the following steps:

  1. Enter service mode in the ACQUITY Console.
  2. In the Sample Manager section of the Console, choose the Maintain pull-down menu, select "Calibrate Rotary Tray", and follow the prompts.
  3. Remove plates and click Next.
  4. Insert the long side of the calibration pin (part number 405013532; usually located in the foam insulation in the Sample Manager door) in the calibration hole and click Next.
  5. Wait for the calibration value to appear, and then click Finish.

ADDITIONAL INFORMATION


Additional information can be found in the video: Calibrating the Rotary Tray in the video library.

Not able to find a solution? Click here to request help.