Skip to main content
Waters

APC System produces higher pressure than expected- WKB59459

Article number: 59459

SYMPTOMS

  • APC System produces higher pressure than expected
  • Pump h/w over pressure error when Sample Manager is in inject mode
  • Pressure increases each time injection is made

ENVIRONMENT

  • SM-pFTN

CAUSE

The seat port 0.003 ID tubing was clogged. By setting a flow and loosening and tightening the fittings along the flow path, it was determined that the high pressure was coming from the sample manager. The needle was removed and solvent was manually flushed through without issue. The seat port tubing was clogged; solvent could not be manually flushed through the tubing.

FIX or WORKAROUND

Replace the seat port 0.003 ID tubing (700006056).

ADDITIONAL INFORMATION

The customer was using preslit Waters vials. Dust from recent construction in the lab may have settled in the customer's sample vials, which were left open on the sample prep bench.

Not able to find a solution? Click here to request help.