Pressure is low on 1525 pump - WKB59268

Article number: 59268

SYMPTOMS

  • Pressure is lower than expected
  • The measured flow rate is lower than the set point

ENVIRONMENT

  • 515 HPLC Pump
  • 1515 Isocratic HPLC Pump
  • 1525 Binary HPLC Pump
  • 1525 Micro Binary HPLC Pump

CAUSE

One or more of the check valves is not sealing properly

FIX or WORKAROUND

Replace the faulty check valve or valves.

ADDITIONAL INFORMATION

How do I change a check valve on my 1500 series pump? - WKB59769