After a recent repair, nanoACQUITY injection sequence is not running the trapping step - WKB5890

Article number: 5890

SYMPTOMS

  • Injection sequence is not running the trapping step

ENVIRONMENT

  • nanoACQUITY
  • MassLynx 4.1

CAUSE

The nanoACQUITY UPLC system/fluidics configuration is set incorrectly to Direct Injection.

FIX or WORKAROUND

  1. Open the ACQUITY UPLC Console.
  2. Highlight the nanoACQUITY UPLC System (top item) in the menu tree on the left-hand side.
  3. Go to Configure > Configure Instrument Modules.
  4. Select Single Pump Trapping.

ADDITIONAL INFORMATION