ACQUITY QSM, QSM-R retention time shifts after system prime - WKB58099

Article number: 58099

SYMPTOMS

  • Retention time initially shifts to a longer retention time than expected after refreshing solvents and repriming the pump.
  • Retention time gradually reverts back to expected retention time after several days and then remains constant until the solvents are refreshed again.

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY Arc

CAUSE

Air within pump

FIX or WORKAROUND

Prime all lines with IPA or methanol for 3 minutes, and then reprime with mobile phases.

ADDITIONAL INFORMATION

 

id58099, A-30SDM, eluent, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, isopropanol, MLYNX, MLYNXV41, SUP, SUPMM, UPAQSM, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARQSM, UPARSMFTN, UPPDAARCB, UPPQSM, UPQSM, UPQSM, UPQSM+, UPQSM+, UPQSMBIO, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO