ACQUITY QSM, QSM-R retention time shifts after system prime - WKB58099
Article number: 58099
SYMPTOMS
- Retention time initially shifts to a longer retention time than expected after refreshing solvents and repriming the pump.
- Retention time gradually reverts back to expected retention time after several days and then remains constant until the solvents are refreshed again.
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY Arc
CAUSE
Air within pump
FIX or WORKAROUND
Prime all lines with IPA or methanol for 3 minutes, and then reprime with mobile phases.