Article number: 5781

SYMPTOMS

  • Error: Z-axis not calibrated

ENVIRONMENT

  • ACQUITY UPLC

CAUSE

The sample manager battery power on CPU is low.

FIX or WORKAROUND

  1. If the error occurs again after calibration, replace the battery.
  • The silver coin 3V battery: part number WAT080443
  • The yellow square 3V battery: part number 700001834
    • Use the dip extractor tool to remove this, part number WAT010339

2. Calibrate the z axis

ADDITIONAL INFORMATION

Check the serial number of the instrument in console and if it has lost the serial number , enter the serial number after replacing battery .

Not able to find a solution? Click here to request help.