When the temperature is set to 4 °C on my ACQUITY H-Class FTN Sample Manager, the display only gets down to 6.2 °C - WKB56884
Article number: 56884
SYMPTOMS
- The display only gets down to 6.2 °C
- The display never reaches the 4 °C set point
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC Sample Manager - Flow Through Needle
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
The sample compartment has a settable range from 4.0 to 40.0 °C with a tolerance range between -2 and +4 °C.
When you set it to 4 °C, it could get to between 2 and 8 degrees and still be within the specifications. (This variation assumes an ambient temperature of less than 23 °C.)
FIX or WORKAROUND
There is no fix. The instrument is working properly and within the specifications.
ADDITIONAL INFORMATION
The temperature accuracy is +/- 0.5 °C. When your temperature stabilizes, it will measure +/- 0.5 °C from the temperature displayed on the instrument (not the set temperature).