2695 system overpressure - WKB56518
Article number: 56518
SYMPTOMS
- System overpressure
- Accumulator overpressure
- Detector flow cell is ok
- Column is ok
- Mobile phases are ok
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2695D Dissolution Separations Module
- 2795 Separations Module
CAUSE
Partial clog in in-line filter
FIX or WORKAROUND
- Wet prime the Alliance at 7.5ml/min for 1 min.
- Back pressure for any mobile phase should be between 100-400psi.
- If pressure is greater than 400psi then likely there is a partial blockage at the inline filter.
- Replace the IN-LINE FILTER ASSY (WAT035190).
- Verify that no further overpressure occurs.
ADDITIONAL INFORMATION
id56518, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, eluent
