Alliance 2695 fails static leak test - WKB55402

Article number: 55402

SYMPTOMS

  • Static leak test fails
  • Parts from the PM kit are not correcting the failure
  • Seals are pressurized

ENVIRONMENT

  • Alliance 2695

CAUSE

The tubing/fittings from the GPV to the primary pump head are faulty.

FIX or WORKAROUND

  1. Change the defective tubing/fittings.
  2. Perform the leak test again.

ADDITIONAL INFORMATION

The part number for the 2695 Standard Flow Gradient Proportioning Valve inlet tubing is WAT270986.