Needle seal readiness test fails on ACQUITY H-Class - WKB5500

Article number: 5500

SYMPTOMS

  • Needle seal readiness test fails

ENVIRONMENT

  • ACQUITY H-Class
  • SM-FTN

CAUSE

System pressure did not stabilize before test or was too low.

A leak is present on the system causing a high pressure difference value.

FIX or WORKAROUND

Set flow should be similar to initial testing (0.6ml/min) with similar pressure (~6000 - 8000psi) and  allow pressure to stabilize before running this test.

If the test continues to fail replace injector parts (Seal, needle and seat port) and re-run the test.

ADDITIONAL INFORMATION

Seat, Vespel w anti-rot, PN 405011492

Assy, Needle 15uL w Guide and Seat, PN 700005215

Seat Port Tubing 0.004 ID, PN 700005234