ACQUITY ELSD has no peaks in trace - WKB5424

Article number: 5424

SYMPTOMS

  • Signal is acceptable in other detector
  • Reference energy of the lamp is >100 nA

ENVIRONMENT

  • ACQUITY ELSD
  • MassLynx

CAUSE

N2 gas not turned on.

FIX or WORKAROUND

  1. Confirm that the N2 is connected to the back fo the detector, and that the regulator is set for 65 - 100 psi.
  2. Turn on N2 in the console.

ADDITIONAL INFORMATION