ACQUITY ELSD has no peaks in trace - WKB5424
Article number: 5424
SYMPTOMS
- Signal is acceptable in other detector
- Reference energy of the lamp is >100 nA
ENVIRONMENT
- ACQUITY ELSD
- MassLynx
CAUSE
N2 gas not turned on.
FIX or WORKAROUND
- Confirm that the N2 is connected to the back fo the detector, and that the regulator is set for 65 - 100 psi.
- Turn on N2 in the console.
ADDITIONAL INFORMATION
id5424, MLYNX, MLYNXV41, SUPMM