Error with 2475: Grating initialization failure, excitation backlash too high - WKB52361

Article number: 52361

SYMPTOMS

  • Error: Grating initialization failure, excitation backlash too high

ENVIRONMENT

  • 2475 Multi-Wavelength Fluorescence Detector

CAUSE

Dirty flow cell

FIX or WORKAROUND

  1. Clean the flow cell as instructed in the operator's manual.
  2. Start a flow.
  3. Reboot the 2475.

ADDITIONAL INFORMATION

 

id52361, 2475, 2475B, A-10FL