How to calibrate the needle on an ACQUITY Sample Manager FTN - WKB51730

Article number: 51730

OBJECTIVE or GOAL

Calibrate the needle z axis.

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class
  • ACQUITY Arc Sample Manager FTN-R

PROCEDURE

Tip: Refer to the diagnostics calibration video to help you complete this task.

  1. Remove the trays and place a business card on the sample tray holder.
  2. From the console, select Maintain > Calibrate needle.
  3. In the Calibrate Needle dialog box, click Start, and then click OK.
  4. Use the +Z button to drive the needle down to within 1 mm of the tray surface.
  5. Switch the displacement increment to 0.1 mm and lower the needle until it just barely touches the business card.
  6. Click Calibrate, and then click Close.

ADDITIONAL INFORMATION

Diagnostics: Calibrate Needle