ACQUITY UPLC Sample Organizer gives "Air Pump time-out" error - WKB51422

Article number: 51422

SYMPTOMS

  • Sample Organizer gives "Air Pump time-out" error
  • Resetting the Sample Organizer clears the error for a few minutes, but it eventually returns
  • Compressor remains on for a while before the error returns

ENVIRONMENT

  • ACQUITY UPLC Sample Organizer

CAUSE

Leak on one pneumatic valve, which is not available as a spare

FIX or WORKAROUND

  1. Replace the entire Pneumatic Panel (700002613). See the procedure Replacing the Pneumatic Panel (715002719).