Skip to main content
Waters

HPLC with 2489 peak tailing in Chromatogram - WKB49549

Article number: 49549

SYMPTOMS

  • SQT Test all passed.
  • Compression test passed
  • Valve test passed  
  • Peak tailing observed in customer sample (black spectra) and same sample, column and solvent used on different instrument (blue spectra) no tailing

peaktailing.JPG

ENVIRONMENT

  • MassLynx
  • Empower
  • 2489 UV/Vis Detector

CAUSE

Bad Flow Cell

FIX or WORKAROUND

Replace with a new flow cell

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.