ACQUITY UPLC BSM or QSM pressure fluctuates - WKB4833

Article number: 4833

SYMPTOMS

Any or all of the following symptoms may be present:

  • ACQUITY BSM or QSM pressure fluctuates
  • Pressure dropout
  • Fails dynamic leak test and Static Leak test
  • Mobile phase coming from the seal wash circuit

    Driplocation.png

ENVIRONMENT

  • Binary Solvent Manager (BSM)
  • Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class
  • ACQUITY PREMIER

clipboard_ea22c01a3ee6c9b21ba94bf72d1eb3afa.png

CAUSE

Leak in the pumphead (seals or checkvalves)

FIX or WORKAROUND

  1. Replace high pressure seals (Part number 700002599 for H-Class pump / 700009135 for I-Class BSM / 700005418 for Premier pumps) and seal wash housing seals (Part number 700002598 for H-Class pump / 700006048 for I-Class BSM / 700005422 for Premier pumps) in the pump head.
  2. Replace Plunger (Part number 700002600 for H-Class pump, I-Class BSM and Premier pumps)
  3. Replace the check valve:

Part numbers for classic ACQUITY:
- primary check valve: 700002596
- primary i2v check valve: 700003563
- accumulator check valve: 700002968

Part numbers for I-Class ACQUITY:
- primary check valve: 700004139
- accumulator check valve: 700003755

Part numbers for H-Class ACQUITY:
- primary check valve: 700005165
- accumulator check valve: 700005164

Part numbers for ACQUITY Premier:
- primary check valve: 700005414
- accumulator check valve: 700010642

clipboard_e5f7a5b0ef06c454594d96e04816dca25.png

ADDITIONAL INFORMATION

For UPLC pump rebuild kit part numbers, see the article Are the primary and accumulator pump rebuild kits for the ACQUITY BSM for two pump heads or one? - WKB25578

This issue can also be caused by a leak in the connector from transducer to check valves