Alliance HPLC System - Solvent leaks onto sample vials- WKB47043

Article number: 47043

SYMPTOMS

Solvent leaks onto the top of sample vials 

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2695D Dissolution Separations Module
  • e2695 Separations Module
  • e2695D Dissolution Separations Module

CAUSE

Worn or torn tubing at the connection of the sample loop to the injector assembly

FIX or WORKAROUND

Replace the sample loop SS Tubing Sample Loop,

  • Phase I, 100 uL, Standard (PN WAT270981),
  • Phase II, 100 µL (PN 430001194),
  • Phase II, 200 µL (PN 430001630),
  • or tighten the fitting

ADDITIONAL INFORMATION

 

id47043, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB