Alliance HPLC System - Solvent leaks onto sample vials- WKB47043
Article number: 47043
SYMPTOMS
Solvent leaks onto the top of sample vials
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2695D Dissolution Separations Module
- e2695 Separations Module
- e2695D Dissolution Separations Module
CAUSE
Worn or torn tubing at the connection of the sample loop to the injector assembly
FIX or WORKAROUND
Replace the sample loop SS Tubing Sample Loop,
- Phase I, 100 uL, Standard (PN WAT270981),
- Phase II, 100 µL (PN 430001194),
- Phase II, 200 µL (PN 430001630),
- or tighten the fitting
ADDITIONAL INFORMATION
id47043, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB