Wavelength accuracy test fails - WKB46657

Article number: 46657

SYMPTOMS

  • Wavelength accuracy fails with a caffeine max of 208 nm
  • Fails after replacing the lamp
  • Flow cell was flushed with acetonitrile and then water with 1% formic acid but continued to fail

ENVIRONMENT

  • ACQUITY UPLC Tunable UV Detector (TUV)
  • ACQUITY UPLC PDA Detector

CAUSE

Dirty flow cell

FIX or WORKAROUND

Repeat the wash for a longer period.

If repeat wash does not work. Replace flow cell.

ADDITIONAL INFORMATION

If the unit continues to fail, perform a 30% phosphoric acid wash.