With 2414 detector, no sensitivity adjustments in RIU mode - WKB45427
Article number: 45427
SYMPTOMS
- In 410 mode, the sensitivity adjustment is available in the method
- In RIU mode, this option is not available
ENVIRONMENT
- 2414 RI detector, Ethernet-controlled
- Empower
CAUSE
Empower autoscales when in RIU mode.
FIX or WORKAROUND
Use 410 mode if sensitivity adjustment is required.