Is there a way to specify a temperature for the nebulizer for the 2424 detector? - WKB45183
Article number: 45183
ENVIRONMENT
- 2424 Evaporative Light Scattering Detector ( ELS)
- ACQUITY UPLC ELS Detector
ANSWER
No.
ADDITIONAL INFORMATION
- You have to specify a power level. The range is 0 to 100%.
- Each percentage is equal to 0.6 °C so 50% is 30 °C.
Note: Only use heating for normal phase solvents.