Error: Leak detected on ACQUITY UPLC BSM - WKB4466
Article number: 4466
SYMPTOMS
- "Leak detected" error message appears when priming the BSM
- Normal flow does not generate the error
ENVIRONMENT
- ACQUITY UPLC BSM
CAUSE
Loose fitting on the vent valve
FIX or WORKAROUND
- Tighten the fittings on the vent valve (port 1 for the solvent B vent line and port 4 for the solvent A vent line).
- Dry the leak sensor.