High delta psi (pressure unstable) on 2695 - WKB44503

Article number: 44503

SYMPTOMS

  • High delta psi (more than 65) when the flow is set to 1 mL/min

ENVIRONMENT

  • e2695
  • 2695

CAUSE

Faulty check valves

FIX or WORKAROUND

  1. Replace the check valves with new valves of the correct type.

ADDITIONAL INFORMATION

 