Skip to main content
Waters

Error: Filter Initialization Failure on 2487, 2489, 2475 or Acquity TUV - WKB4443

Article number: 4443

SYMPTOMS

  • Error message: Filter initialization failure

ENVIRONMENT

  • 2487 Dual-Wavelength Absorbance Detector
  • 2489 UV/Vis Detector
  • 2475 Multi-Wavelength Fluorescence Detector
  • ACQUITY UPLC Tunable UV Detector

CAUSE

Bubble in the flow cell

FIX or WORKAROUND

  1. Flush the flow cell with solvent at 0.5 ml/min for at least 15 minutes.
  2. Reboot the detector after 15 minutes with flow on and let it initialize
  3. Do a wavelength calibration as per detector manual

ADDITIONAL INFORMATION

If 100% Acetonitrile does not work, try with 100% MeOH instead

id4443, 2475, 2475B, 2487, 2489, 2489B, 2489SFC, A-10FL, A-10UV, A-30UV, methanol, UPARPTUV, UPPTUV, UPTUV, UPTUVARC, UPTUVARCB, UPTUV-E, UPTUVTC

Not able to find a solution? Click here to request help.