Error: Filter Initialization Failure on 2487, 2489 or 2475 - WKB4443
Article number: 4443
SYMPTOMS
- Error message: Filter initialization failure
ENVIRONMENT
- 2487 Dual-Wavelength Absorbance Detector
- 2489 UV/Vis Detector
- 2475 Multi-Wavelength Fluorescence Detector
- TUV
CAUSE
Bubble in the flow cell
FIX or WORKAROUND
- Flush the flow cell with solvent at 0.5 ml/min for at least 15 minutes.
- Reboot the detector after 15 minutes with flow on and let it initialize
- Do a wavelength calibration as per detector manual
ADDITIONAL INFORMATION
id4443, 2475, 2475B, 2487, 2489, 2489B, 2489SFC, A-10FL, A-10UV