Sample Manager FTN "Washing Needle" message - WKB430

Article number: 430

SYMPTOMS

The Sample Manager FTN will not proceed with an injection and displays a "Washing Needle" message in Console.

ENVIRONMENT

  • H-Class FTN
  • H-Class BioFTN
  • I-Class FTN

CAUSE

An external signal cable connected to the "Hold" input on the rear panel I/O connection triggers this message.

FIX or WORKAROUND

Ensure that there is no signal cable connected to the "Hold" input on the rear panel I/O connection.

ADDITIONAL INFORMATION