Sample Manager FTN "Washing Needle" message - WKB430
Article number: 430
SYMPTOMS
The Sample Manager FTN will not proceed with an injection and displays a "Washing Needle" message in Console.
ENVIRONMENT
- H-Class FTN
- H-Class BioFTN
- I-Class FTN
CAUSE
An external signal cable connected to the "Hold" input on the rear panel I/O connection triggers this message.
FIX or WORKAROUND
Ensure that there is no signal cable connected to the "Hold" input on the rear panel I/O connection.
ADDITIONAL INFORMATION
id430, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO