ACQUITY UPLC Sample Manager (SM) needle often crashes - WKB4203

Article number: 4203

SYMPTOMS

  • Needle damaged

  • XYZ crash or damage

  • Zp crash or damage

ENVIRONMENT

  • ACQUITY UPLC Sample Manager (SM)

CAUSE

  • Misalignment of the needle

FIX or WORKAROUND

  1. Check the needle placement and the vials.
  2. MassLynx or Empower - adjust as described on page 11 of Sample Vials and Accessories (720001818).

ADDITIONAL INFORMATION