How to change the IEEE address on a 717 autosampler ? - WKB4129
Article number: 4129
OBJECTIVE or GOAL
How to change/reset the IEEE address on a 717 autosampler.
ENVIRONMENT
- 717 plus Autosampler
PROCEDURE
- Verify that the address is a unique number from 2 to 29. The default IEEE-488 address is set to 30.
- From the 717+ Main Menu, press Config Page.
- From Configuration Page 1, press Next Cfg Page three times to view Page 4 of the Configuration Pages.
- On Configuration Page 4, highlight the desired IEEE address field, type the new address, and press Enter (see the following screen shot).
- Press Reboot System to configure and store the new address.
ADDITIONAL INFORMATION
id4129, 717, 717HC, 717PLUS, 717UP