How to change the IEEE address on a 717 autosampler ? - WKB4129

Article number: 4129

OBJECTIVE or GOAL

How to change/reset the IEEE address on a 717 autosampler.

ENVIRONMENT

  • 717 plus Autosampler

PROCEDURE

  1. Verify that the address is a unique number from 2 to 29. The default IEEE-488 address is set to 30.
  2. From the 717+ Main Menu, press Config Page.
  3. From Configuration Page 1, press Next Cfg Page three times to view Page 4 of the Configuration Pages.
  4. On Configuration Page 4, highlight the desired IEEE address field, type the new address, and press Enter (see the following screen shot).
  5. Press Reboot System to configure and store the new address.

Setting IEEE address on 717 .PNG

ADDITIONAL INFORMATION