ACQUITY Binary Solvent Manager (BSM) continuous beep on power-up - WKB4067

Article number: 4067

SYMPTOMS

  • ACQUITY UPLC Solvent Manager (BSM or QSM) continuous beep on power-up.

ENVIRONMENT

  • ACQUITY UPLC Binary Solvent Manager (BSM)
  • ACQUITY UPLC Quaternary Solvent Manager (QSM)
  • NanoAcquity
  • Auxillary Nano Solvent Manager

CAUSE

The battery is flat (low current).

FIX or WORKAROUND

Replace the battery on both the Pump A and Pump B CPU PCBs.

  • 3vCCU PCB uses coin type "Battery, 3V Lithium" (WAT080443)
  • CPU 2000 uses yellow square "Battery, 3V Zeropower, Snaphat"(700001834)

ADDITIONAL INFORMATION

 