How to remove the collet from a compressed reusable high-pressure fingertight fitting - WKB3756
Article number: 3756
OBJECTIVE or GOAL
Remove the collet from a compressed reusable high-pressure fingertight fitting.
ENVIRONMENT
- ACQUITY column stabilizer
PROCEDURE
1. Order Part Number 700003170 for the "MULTI-TOOL, COLLET & COMP. SCREW, HPFT".
2. Use the tool to remove the collet from the compression screw.
3. Replace the PEEK ferrule (700003114 for the FERRULE, PEEK, 1/16, HPFT, PKG 10).
ADDITIONAL INFORMATION
Part Number compression screw 700003169 (10 pack)
Part Number for collet 700003168 (10 pack)
id3756, UPBINARY, UPCMIVD, UPCOLMGR