Skip to main content
Waters

Cannot inject from plate 3 or higher with the ACQUITY Sample Organizer - WKB3623

Article number: 3623

SYMPTOMS

  • When attempting to inject from a plate higher than 2, a bad plate location error message appears

ENVIRONMENT

  • MassLynx 4.1
  • ACQUITY Sample Organizer

CAUSE

The Bed Layout is set up for a Sample Manager with two plates, not for a Sample Organizer with 10 or so plates.

FIX or WORKAROUND

  1. To configure the Inlet Editor for more than two plates, go to the ACQUITY Sampler menu in the Inlet Editor and select Bed Layout.

Acquity Sampler.PNG

  1. Type the new name (for example, 10 Plate Bed) in the top white space under "Bed Layouts:".

10 Plate Bed Layout.PNG

  1. In the small white field to the right of where you typed the name, type in the total number of plates (including tray 1, which sits on the right in the sample manger) and hit the column.PNG button. Click elsewhere and the number typed should remain.
  2. Select the save new bed layout button Save New Bed.PNG.
  3. Go to Current Layout Selected and select the new bed layout name from the drop-down menu. 
  4. Select OK.
  5. In the inlet editor, select the LC drop-down menu and select Reset Communications.
  6. Run a test injection to make sure that you can now select from more than two plates. 

ADDITIONAL INFORMATION

 

id3623, comms, MLYNX, MLYNXV41, SUPMM

Not able to find a solution? Click here to request help.