Method editor crashes when switching from direct inject to trapping mode in Analyst and Xcalibur with an ACQUITY M-Class - WKB32323
Article number: 32323
SYMPTOMS
- Direct inject works properly
- When switching from direct inject to trapping mode, the method editor crashes with no error message
- Or unable to run a trapping method
ENVIRONMENT
- Analyst
- ACQUITY M-Class
- nanoACQUITY UPLC
- Xcalibur
- M-Class/nanoACQUITY Driver Pack 1.55
CAUSE
Incorrect folder permissions
FIX or WORKAROUND
- Browse to the Methods folder (/program files (x86)/waters instruments/methods for Analyst) or where customer stores methods in Xcalibur
- Right-click on the Methods folder and select Properties
- On the Security tab, set read/write access for all users
- Click OK to save the changes and close out of Properties
- Reopen method editor and confirm trapping can be configured properly