Extended needle wash may cause area variation with Alliance HPLC - WKB306894
Article number: 306894
SYMPTOMS
- Area variation
- RSD failure
- Low response
ENVIRONMENT
- Alliance HPLC
- 2690/5 HPLC
- Alliance with 3.04 firmware
CAUSE
Selection of Extended Needle Wash mode selection, Run needle wash both pre and post while sample is in place at needle may cause area count variation.
FIX or WORKAROUND
- Do not use needle wash mode as EXTENDED in Alliance HPLC having firmware 3.04.
- Always use needle wash mode as NORMAL or Double with this firmware.
ADDITIONAL INFORMATION
id306894, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB