Extended needle wash may cause area variation with Alliance HPLC - WKB306894

Article number: 306894

SYMPTOMS

  • Area variation
  • RSD failure
  • Low response 

ENVIRONMENT

  • Alliance HPLC
  • 2690/5 HPLC
  • Alliance with 3.04 firmware

CAUSE

Extended Needle Wash mode selection, Run needle wash as both pre and post while sample is in place at needle. This may cause area count variation. 

FIX or WORKAROUND

  1. Do not use needle wash mode as EXTENDED in Alliance HPLC having firmware 3.04.
  2. Always use needle wash mode as NORMAL or Double with this firmware.

ADDITIONAL INFORMATION