Injection reproducibility issue with Alliance and Empower - WKB30289

Article number: 30289

SYMPTOMS

  • Peaks are much smaller than usual (low height and area)
  • Column is not an issue
  • Temperature normal

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2695D Dissolution Separations Module
  • 2489 UV detector
  • Empower 3
  • Aqueous buffer mobile phase 

CAUSE

Possibly dirty flowcell, or injector has bubbles.

FIX or WORKAROUND

  1. Purge the injector for 20 cycles.
  2. Run the needle wash/seal wash prime for 10 minutes (Home > Diag). 
  3. Flush the system with 100% HPLC-grade water to remove any salt buildup.
  4. Wet prime with mobile phase.
  5. Repeat run.

ADDITIONAL INFORMATION

 