What position should the needle be in for a nanoACQUITY? - WKB30179
Article number: 30179
ENVIRONMENT
- nanoACQUITY
- ACQUITY M-Class
ANSWER
For ICS 1.42 and earlier:
- The needle should be in position 2 of the injection valve.
- The bubble detect (VDD) should be in position 3 of the injection valve.
For ICS 1.53 and later:
- The needle should be in position 3 of the injection valve.
- The bubble detect (VDD) should be in position 2 of the injection valve.
ADDITIONAL INFORMATION
If the needle or the VDD had to be moved, prime the syringes, wash the needle, and then characterize the needle and loop volumes.