Agilent GC 8890 pressure issue - WKB300842

Article number: 300842

SYMPTOMS

  • Agilent GC 8890 pressure issue
  • Pressure not stable

ENVIRONMENT

  • Empower 3 Personal
  • Agilent 8890 GC

CAUSE

Defective Analog valve connected on inlet EPC and located on split vent trap housing

FIX or WORKAROUND

Replace the Analog valve (part G3450-80019).

ADDITIONAL INFORMATION