How do I connect PEEK tubing to the PDA inlet using a PEEK compression screw and ferrule? - WKB30014
Article number: 30014
OBJECTIVE or GOAL
Connect PEEK tubing to the PDA inlet using a PEEK compression screw and ferrule.
ENVIRONMENT
- Waters PDA instrument, all models
PROCEDURE
- Slide the PEEK compression screw and ferrule onto the end of the tubing.
- With one hand, push the tubing into the PDA inlet as far as it will go.
- Hold the tubing in the inlet, keeping pressure on it so that it doesn't back out at all.
- With the other hand, tighten down the compression screw.
- This seats the ferrule in the inlet, at the correct distance from the end of the tubing.
ADDITIONAL INFORMATION
id30014, A-30PDA, UPAPDA, UPC2PDA, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA