How to replace the injection valve on an SM-FL - WKB298517
OBJECTIVE or GOAL
Replace the injection valve on an SM-FL.
ENVIRONMENT
- ACQUITY UPLC I-Class SM-FL
- ACQUITY UPC2 System
PROCEDURE
1. In the ACQUITY Console, select Sample Manager FL from the system tree.
2. In the SM-FL information window, click Maintain > Park needle and inject valve > Yes.
3. Power-off the SM-FL.
4. Open the fluidics compartment door.
5. Remove the fittings attached to the injection valve cartridge.

6. Use a 2-mm hex wrench to remove the screw at the 10 o’clock position on the injection valve cartridge.
7. Remove the injection valve cartridge from the injection valve assembly.
8. Carefully unpack the replacement injection valve cartridge.
9. Ensure that the groove in the cartridge housing and the groove on the drive clamp align.
Tip: If they do not align, turn the drive clamp until the grooves line up.

10. Insert the new injection valve cartridge into the injection valve assembly.
Note: If the injection valve cartridge does not slide fully into the injection valve assembly, contact your Waters service representative.
11. Insert the 2-mm hex screw at the 10 o’clock position on the injection valve cartridge, and tighten it.
12. Reattach all fittings, ensuring that the tubes bottom in their fitting holes.
13. Close the fluidics compartment door.
14. Power-on the SM-FL.
15. Characterize the sample needle and sample loop volumes.