ELS Detector gives poor signal or flat-line data - WKB297874

Article number: 297874

SYMPTOMS

  • ELSD signal is flat 
  • ELSD signal is inconsistent

ENVIRONMENT

  • ACQUITY UPLC ELS Detector
  • 2424 Evaporative Light Scattering Detector
  • 2420 Evaporative Light Scattering Detector

CAUSE

Dry drain tubing leaves ELSD drift chamber unsealed. There must be liquid draining or else the signal will not be stable.

FIX or WORKAROUND

Use a syringe or water bottle to add water or other liquid to the J drain tubing below the nebulizer.

ADDITIONAL INFORMATION