GPV (Composition Accuracy) test failure with propylparaben mobile phase in 2695/Alliance - WKB297303
Article number: 297303
SYMPTOMS
- GPV test failure with A/B/C and A/B/D lower than 9.5 %
ENVIRONMENT
- 2690/5
- Alliance
CAUSE
- Check valve failure
- C/D ratio higher than 1.60
FIX or WORKAROUND
- Inspect Diag for a C/D ratio lower than 1.60.
- Sonicate the check valve in 100% water and then 100% methanol.
- If the C/D value is still high, replace the check valve.
- 700000254 - Ruby Standard Check Valve
- 700002399 - Ceramic Check Valve
ADDITIONAL INFORMATION
During GPV test, pressure will be constant. Delta psi will be within a 2.5% limit of system pressure.
