"Drawing sample rate excessive" occurs in H-Class FTN frequently. - WKB29359
Article number: 29359
SYMPTOMS
- “Drawing sample rate excessive” error occurs frequently
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- 186005837 96-well Sample Collection Plate, 700uL Round well
http://www.waters.com/waters/partDetail.htm?partNumber=186005837 - 186006336 Adhesive Seal for 96 and 384 well plates
http://www.waters.com/waters/partDetail.htm?partNumber=186006336
CAUSE
186006336 plate seal is blocked
FIX or WORKAROUND
Use 186006332 Cap-mat 96 well 7 mm round plug pre-slit silicone/PTFE.
http://www.waters.com/waters/partDetail.htm?partNumber=186006332
ADDITIONAL INFORMATION
id29359, A-30SM, UPISMFTN, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO