"End of run timed out" error on 2414 Refractive Index Detector - WKB293383

Article number: 293383

SYMPTOMS

  • "Instrument failure - End of run timed out at xxxx.x seconds. No reply from W2414#Serial Number" in message center
  • Does not occur when injecting a sample/standard, only with Purge detector, Condition column, and Equilibrate

ENVIRONMENT

  • Empower 3
  • 2414 Refractive Index Detector
  • RIU Mode

CAUSE

The 2414 method type of the instrument method used for purge/condition/equilibration is set to Acquisition.

FIX or WORKAROUND

  1. For the instrument method used for stabilization, modify the method type from Acquisition to Purge.
  2. When in W410 mode, the device is set according to the function of the sample set. However, in RIU mode, you must set the instrument method according to the function.